en Caret down
Contacts: Phone icon +380679996889 sss-labor@ukr.net

Description

Intended for the determination of all types of oils and fats of animal and vegetable origin:

- acidity (FFA),

- peroxide number (PV),

- p-Anisidine number (AnV),

-          dear,

- polyphenols (OSI),

- iodine number,

The result is automatically calculated, displayed and printed.

It has high sensitivity, a wide measurement range and excellent repeatability of results thanks to innovative photometric technology that uses LED light sources and a fixed wavelength in the range from ultraviolet to visible spectrum (with an optical density range from 0 to 6).

The results of the analysis are correlated with the results of control methods.

Extracted fats

 

Analysis

Measuring range

resolution

Frequency

Analysis time

Acidity (FFA)

 

0.01 - 1.10% oleic acid

1.0 – 3.5% oleic acid

1.0 – 26.0% oleic acid

 

 

0.01% oleic acid

0.1% oleic acid

0.1% oleic acid

 

 

0.02% oleic acid

0.1% oleic acid

0.5% oleic acid

 

1 min

1min

1 min

Peroxide number

0.01 – 5.50 meqO²/Kg

0.30 – 25.00 meqO²/Kg

1.0 – 50.0 meqO²/Kg

4.0 – 550.0 meqO²/Kg

0.01 meqO²/Kg

0.01 meqO²/Kg

0.1 meqO²/Kg

0.1 meqO²/Kg

0.05 meqO²/Kg

0.24 meqO²/Kg

0.5 meqO²/Kg

5.2 meqO²/Kg

3 min

3 min

3min

3 min

p-Anisidine number

0.5 – 100.0 А.h.

0.1 А.h.

0.7 А.h.

1 min

Soap

1 – 300 ppm

50 – 1350 ppm

100 – 3000 ppm

1 ppm

1 ppm

1 ppm

9 ppm

38 ppm

77 ppm

1 min

1 min

1 min

Iodine number (in palm, coconut and kernel oil)

2 – 100 I.h.

1 I.h.

 2 I.h

1min

 

Polyphenols

(for olive oil)

 

200 – 1000 mg/kg Tyrosol

1 mg/kg Tyrosol

34 mg/kg Tyrosol

5 min

Oil Stability Index (OSI)

9.0-40.0 h.

0.1 h.

0.8 h.

 

Analysis of oils and fats - Full analysis panel

Samples that can be analyzed simultaneously - 16

Multitasking mode – Yes

Calibration - Pre-calibrated. No need for periodic calibration

Maintenance costs – None

Storage of results - Internal memory for saving thousands of analysis results in CSV and XML files, compatible with all database formats (eg: XLS, SQL)

Photometric module - up to 8 wavelengths in 4 reading cells

Incubation module - 16 thermostated cells with a temperature of 37°C

Connection to scanners of barcodes and QR codes - Yes, via Bluetooth

Display - 5.7-inch color TFT LCD with touch screen

Connectivity - 1 USB Type B port for transfer of analysis results database, configuration and software update, PC connection 1 USB Type A port for maintenance and PC connection 1 Ethernet (LAN) port for internal network connection Bluetooth 4.0

Printer - 80 mm wide printer with built-in graphics

Dimensions and weight - 32 x 29.5 x 13 cm (W x D x H) 2.80 kg

Power supply unit - 24 V