Description
Uses built-in computer technology to control 8 optical channels and features such as automatic measurement, analysis and data processing.
The reader has four standard spectral filters, can perform automatic identification of single- or double-wavelength measurements according to the settings.
The instrument has a large LCD display and touch screen.
It can be used for diagnostics of various clinical studies, immunology, microbiology, veterinary medicine, biochemistry, biochemical genetics, endocrinology, etc.
Measurements of optical absorption: quantitative, semi-quantitative, qualitative modes.
8-channel measurement system.
Ability to read from a single well.
Reader software and developed PC software.
Extensive quality control program with Levy-Jennings curve.
Allows for dual wavelength reading
Optical system provides high accuracy and speed.
Readout time for 96-well microplate is 8 or 18 sec. using single and dual wavelengths respectively.
Sample counting mode: available with several standard non-linear types next to Cone Loge, Loge loge, Multipoint, Linear, O.D log.
Specifications:
Determination principle - vertical photometry
Modes of analysis - semi-quantitative, quantitative, qualitative
Microplate type - 96-well plate
Measuring system - 8-channel fiber optic system
Measuring speed - monochromatically in 8 seconds; biochromatically in 18 seconds.
Light source - halogen 12 V, 20 W bulb
Wavelength - 405, 450, 492, 630 nm (with 3 additional filters)
Display - monochrome LCD STN-BLUE 320x240 pixels with LED backlight
Keypad - touch screen
Measuring range - 0.000 - 4.000 O.D.
Reproducibility - CV ± 0.3%
Stability - CV ± 0.003 O.D. per 1 h
Linearity - 0.000 to 3.000 O.D. for 492 nm
Shaking - 5 modes (from weak to strong), vibration time from 0 to 240 sec
Data transmission - RS-232, external USB-printer
Power - 230 V, 50 Hz (max 45 W)
Operating environment - from 5 ° C to 40 ° C
Dimensions - 460 mm (W) x 380 mm (D) x 370 mm (H)
Weight ~15 kg